AQUALAB TDL

Firmware Updates

Ben Walden
Updated by Ben Walden
For instruments that are user calibrated at temperatures other than 25 °C, new firmware may be needed to support our new potassium acetate verification standards. See the following page for more information.

Update Instructions

Our AQUALAB instruments support firmware updates, allowing you to benefit from the latest features and bug fixes. To update the firmware on your device:

  1. Select your desired firmware version from the table below.
  2. Download the Firmware Updater Application.
  3. Connect your instrument to your computer using a USB or serial cable.
  4. Run the Firmware Updater Application and follow the on-screen instructions.
  5. After a firmware update, it is recommended that you run a water activity verification of your instrument.

Check Your Firmware Version

You can only update your instrument to a new firmware version if it has the same first number in its version (e.g. v2.#.# and v3.#.#)

To check which version your instrument is running:

  1. Navigate to the Config menu on your instrument.
  2. Scroll down to the About screen.
  3. Look for the Version number.

Firmware Change Log

Version

Release Date

Change Log

Firmware Updater Application

3.6.0

09/19/2025

  • Rebranded to Addium
  • Update 0.25 and 0.50 aw verification standards to use potassium acetate temperature corrections.

3.5.11

06/26/2024

  • Fixed bug causing reading duration and continuous mode symbol to be omitted reading screen
  • Fixed rare aw rounding error which resulted in the aw being off by ±0.0001 aw
  • Improved temperature control performance at extreme temperatures, particularly 15°C
  • Updated translations
  • Minor bug fixes

3.5.10

09/14/2023

  • Improved base temperature control.  Now more stable.
  • Updated translations
  • Fix bug causing fast aw readings to be stored with the timestamp "01/01/00, 12:00AM"
  • Fix bug causing inconsistencies between the fast aw results in the device reading history and in SKALA
  • Fix bug that causes a rare rounding error between the screen / serial printer output and the reading history
  • Minor bug fixes

3.4.18

12/08/2022

  • Minor bug fixes

3.4.17

11/28/2022

  • Allow a variable number of verification standards for multi-point calibrations
  • Add option in admin screen to configure default set of multipoint standards
  • Improved laser temp control stability.
  • Minor bug fixes

2.24.1

02/28/2020

  • Fix sample temperature rounding error issue.
  • Final firmware updater support for legacy AQUALAB TDL instrumentation.