AQUALAB 4

Firmware Updates

Ben Walden
Updated by Ben Walden
For instruments that are user calibrated at temperatures other than 25 °C, new firmware may be needed to support our new potassium acetate verification standards. See the following page for more information.

Update Instructions

Our AQUALAB instruments support firmware updates, allowing you to benefit from the latest features and bug fixes. To update the firmware on your device:

  1. Select your desired firmware version from the table below.
  2. Download the Firmware Updater Application.
  3. Connect your instrument to your computer using a USB or serial cable.
  4. Run the Firmware Updater Application and follow the on-screen instructions.
  5. After a firmware update, it is recommended that you run a water activity verification of your instrument.

Check Your Firmware Version

You can only update your instrument to a new firmware version if it has the same first number in its version (e.g. v2.#.# and v3.#.#)

To check which version your instrument is running:

  1. Navigate to the Config menu on your instrument.
  2. Scroll down to the About screen.
  3. Look for the Version number.

Firmware Change Log

Version

Release Date

Change Log

Firmware Updater Application

3.06.0

09/19/2025

  • Rebranded to Addium
  • Updated 0.25 and 0.50 aw verification standards to use potassium acetate temperature corrections.

3.04.10

06/26/2024

  • Fixed bug introduced in v3.04.9 causing reading duration and continuous mode symbol to not appear on the reading screen

3.04.9

11/06/2023

  • Fixed bug causing rare rounding error between the screen and serial printer output where aw was off by ±0.0001 aw

3.04.8

10/06/2021

  • Fixed system issue when coin cell backup battery is missing or low

3.04.7

10/06/2021

  • Added slope and offset water activity correction capabilities
  • Fixed time drift issue when instrument is hard reset.
  • Fixed issue where firmware update locks user offset calibration for cap sensor type.
  • Fixed reading 0 temp 0 issue desiccant readings.
  • Translation updates

3.02.1

02/28/2020

  • Fixed sample temperature rounding error issue.

2.16.0

06/15/2023

  • Fix sample temperature rounding error issue.
  • Final firmware updater support for legacy AquaLab Series 4TE/TEV instrumentation.